Interchannel
Crosstalk Evaluation in a Prototype Phased Array Ultrasound System Designed for
Hyperthermic Cancer Therapy
D.A. Chorman1, R.J.
McGough1, T. Hall2, and C. A. Cain2
1Biomedical
Ultrasonics and Electromagnetics Laboratory
2University of
2Biomedical
Ultrasonics Laboratory
Abstract: Ultrasound
phased array prototypes designed for hyperthermia breast cancer therapy have
been constructed and are being evaluated. Preliminary designs for power
amplification and electronic steering systems used for driving and phasing the
array prototypes have been developed and interfaced with a computer for PC
control. Field programmable gate arrays (FPGA) digitally generate
continuous-wave or pulsed control signals, which are amplified and transmitted
to the ultrasound transducers. The
high-power square-wave outputs of the power amplifiers are filtered using high
gain resonant circuits, and the fundamental sinusoidal component powers each
transducer. The phased array system is tested in a water tank using a
two-dimensional positioning system and a hydrophone. Focal pattern mapping is
automated with a 2-D
hydrophone positioning system that is computer-controlled through a MATLAB
interface.
Array steering is improved with
shape calibration, which is a triangulation procedure that determines the
spatial orientation of the individual array elements with respect to the
coordinate axes defined by the positioning system. Shape calibration results
are degraded by crosstalk between the output power amplifier channels.
Crosstalk excites multiple elements instead of a single transducer, which
produces triangulation errors during the shape calibration. Previous
experiments have shown that the crosstalk voltage between adjacent channels
should be at or below a maximum value of -30dB for satisfactory shape
calibration results.
Crosstalk measurements were
performed by driving a single channel and measuring the voltage induced on
adjacent channels. The voltage measurements were extracted from the outputs of
the power amplifier filters. The driving signal consisted of two cycles of a
971.5 KHz square-wave, which was filtered after amplification. The crosstalk
lasted for 30µs immediately after the initiation of the pulse sequence, and the
peak voltage value was reached in less than 5µs. The crosstalk values were
extracted from peak-to-peak voltages occurring in the 0-30µS time frame. For
one prototype array, the crosstalk values range from -30.59dB to -38.74dB.
Thus, the measured crosstalk values for this prototype ultrasound phased array
are in an acceptable range for shape calibration.
Prototype
Ultrasound System Overview
The block diagram below shows
the components of the prototype ultrasound system. The components are discussed
in more detail below.

Digital Signal Generator:
A
digital signal generator synthesizes the low-level signals that are applied to
the individual channels of the ultrasound system. The signal generator was
constructed using field programmable gate arrays (FPGA’s). The frequency,
phase, and mode (Pulsed or CW) of operation is programmable through a USB
connected PC, using a MATLAB® command line interface. The Digital Signal Generator boards are shown
below in Figure 1.
Figure 1: Digital Signal Generator Boards
Class D Power Amplifier:
The
low-level signals produced by the digital signal generator are amplified by a
Class-D power amplifier. This amplifier can output a maximum of 10 watts in continuous
mode and can operate up to a maximum frequency of 1MHz. More powerful amplifier
designs that will output 2-3X the power and operate at a frequency maximum of
3MHz are also being investigated.
Figure 2: Class D Power Amplifier Board
Resonant
The
square-wave output of the power amplifier is filtered using a resonant low pass
filter. The filter: 1. removes the DC offset component from the output signal; 2.
filters out all of the higher harmonics contained in the square-wave, passing only the fundamental sinusoidal
component; and 3. resonates at the fundamental frequency resulting in a voltage
output amplitude that is 5-10X the input
voltage.
Figure 3: Filter Board
Ultrasound Array
The interconnect is greatly simplified with a flex
circuit interface, resulting in a robust and reliable electrical connection.
The geometry in this prototype system is a 1D linear phased array.
Figure 4: Ultrasound Array with a Flexible Interconnect